Fluorescence Filters

Interference filters are essential optical components for fluorescence measurements in life science and biophotonics. These measurements require a pair of filters: the first filter (exciter filter ) for the transmission of the excitation beam to cause the absorption in the sample and the second filter (emitter filter) for the fluorescent color to pass and meanwhile block the residual excitation energy to get into the detector. Our high performance filter sets are designed for a large variety of fluorophores and have a high peak transmissions of >92% in the pass band and OD6 blocking in the stop band. 

Learn more about thin film filters

Standard Fluorescence Filter Sets

Part # Fluorophore Exciter Emitter
Center
Wavelength
Typical Bandwidth Center Wavelength Typical Bandwidth
FF-DAPI DAPI 365 nm 40 nm 460 nm 50 nm
FF-CFP CFP 436 nm 20 nm 480 nm 30 nm
FF-GFP GFP 470 nm 40 nm 525 nm 50 nm
FF-FITC FITC 480 nm 30 nm 535 nm 40 nm
FF-YFP YFP 500 nm 20 nm 543 nm 30 nm
FF-TRIT TRITC 530 nm 50 nm 605 nm 55 nm
FF-CY3 CY3 545 nm 30 nm 610 nm 75 nm
FF-TEXR Texas RedR 560 nm 40 nm 630 nm 60 nm
FF-CY5 CY5 620 nm 60 nm 695 nm 70 nm

Exciter/Emitter Filter Specifications

Substrate Material: Low fluorescent BK7 glass or equivalent
Substrate Size: Diameter: 10 or 15+0.0/-0.1 mm, Thickness: 3.3 ± 0.1 mm
Holder Size: Diameter: 25 +0.0/-0.1 mm, Thickness: 5.0 ± 0.1 mm
Passband Transmission: >92% (except DAPI)
Reflection Band: N/A
Surface Quality: 60/40 scratch and dig
Clear Aperture: >85% of central area
Blocking: OD6 outside of pass band
 